CHARGEP PORT SAFETY CAP - GB STANDARD (Remove and Replace)
Correction code
4401010702
FRT
0.00
NOTE: Unless
otherwise explicitly stated in the procedure, the above correction code and
FRT reflect all of the work required to perform this procedure, including
the linked procedures. Do
not stack correction codes unless explicitly told to do so.
NOTE: See Flat Rate Times to learn more about FRTs and how they are
created. To provide feedback on FRT values, email LaborTimeFeedback@tesla.com.
NOTE: See Personal Protection to make
sure wearing proper PPE when performing the below procedure.
- 1076921-00-A Insulation Multimeter, Fluke 1587 (EMEA)
- 1130480-00-A Test Probes, Slim, Fluke TP38
Only
technicians who have been trained in High Voltage Awareness and
have completed all required certification courses (if applicable) are permitted to
perform this procedure. Proper personal protective equipment
(PPE) and insulating HV gloves with a minimum rating
of class 0 (1000V) must be worn at all times a high voltage
cable, busbar, or fitting is handled. Refer to Tech Note TN-15-92-003, High Voltage Awareness
Care Points
for additional safety
information.
Remove all jewelry (watches, bracelets, rings, necklaces, earrings, ID tags, piercings, etc.) from your person, and all objects (keys, coins, pens, pencils, tools, fasteners, etc.) from your pockets before performing any procedure that exposes you to high voltage.
Proper Personal Protective Equipment (PPE) is required to perform this procedure:
- High Voltage (HV) insulating gloves
- Leather glove protectors
- High voltage glove tester
- Safety glasses
- Electrical hazard rated safety shoes
A glove inflator is the only recommended way to test HV gloves. Both HV gloves must pass testing before beginning this procedure. If either glove does not pass the air check, discard the pair.
- North America: HV gloves can be used up to 12 months after the testing date printed on the glove, but only 6 months after first use even if the gloves are still within the 12 month period. If the gloves were not put into service during the 12-month period following the testing date, they must be retested before being put into service.
- Europe: Class 0 (1000V) HV gloves have no expiration date and may be used if they pass the pre-use assessment and the periodic inspection.
- Asia Pacific: Follow the
expiration date stated by the HV glove manufacturer, and:
- o China only: Test the gloves every 6 months according to GB/T 17622.
Remove
- Remove the 2nd row lower seat cushion. See 2ND ROW SEAT CUSHION (Remove and Replace).
- Remove the rear underhood apron. See Underhood Apron - Rear (Remove and Replace).
- Perform a Charge Port Voltage Check. See Charge Port Voltage Check.
- Remove the LH trunk load floor. See TRUNK LOAD FLOOR - REAR (Remove and Replace).
- Remove the trunk sill trim. See TRUNK SILL LOWER TRIM (Remove and Replace).
- Remove the LH trunk load floor bracket. See Bracket - Load Floor - LH (Remove and Replace).
- Remove the trunk side trim. See TRUNK SIDE TRIM LEFT HAND - REAR UNDER BODY (Remove and Replace).
- Disconnect the DC logic electrical connector from the charge port assembly.
- Release the tabs (x2) and clip (X1) that attach the charge port safety cap to the charge port assembly, and then remove the safety cap.
Install
- Install the charge port safety cap on the charge port assembly.
- Connect the DC logic electrical connector from the charge port assembly.
- Install the trunk side trim. See TRUNK SIDE TRIM LEFT HAND - REAR UNDER BODY (Remove and Replace).
- Install the LH trunk load floor bracket. See Bracket - Load Floor - LH (Remove and Replace).
- Install the trunk sill trim. See TRUNK SILL LOWER TRIM (Remove and Replace).
- Install the LH trunk load floor. See TRUNK LOAD FLOOR - REAR (Remove and Replace).
- Connect the LV power. See LV Power (Disconnect and Connect).
- Install the 2nd row lower seat cushion. See 2ND ROW SEAT CUSHION (Remove and Replace).
- Install the rear underhood apron. See Underhood Apron - Rear (Remove and Replace).